Spatial and Temporal Analysis of Bias HAST System Temperature

Pfeifer, Kent B.; Furrer III, Clint T.; Sandoval, Paul A.; Garrett, Stephen E.; Pfeifer, Nathaniel B.

High-reliability components for high-consequence systems require detailed testing of operation after having undergone highly accelerated stress testing (HAST) under unusual conditions of high-temperature and humidity. This paper describes the design and operation of a system called "Wormwood" that is a highly multiplexed temperature measurement system that is designed to operate under HAST conditions to allow measurement of the temperature as a function of time and position in a HAST chamber. HAST chambers have single-point temperature measurements that can be traceable to NIST standards. The objective of these "Wormwood" measurements is to verify the uniformity and stability of the remaining volume of the HAST chamber with respect to the single traceable standard.