Publications
X-ray power and yield measurements at the refurbished Z machine
Jones, Brent M.; Ampleford, David A.; Cuneo, M.E.; Hohlfelder, Robert J.; Jennings, C.A.; Johnson, Drew J.; Jones, Brent M.; Lopez, M.R.; Macarthur, J.; Mills, Jerry A.; Preston, T.; Rochau, G.A.; Savage, Mark E.; Spencer, D.; Sinars, Daniel S.; Porter, J.L.
Advancements have been made in the diagnostic techniques to measure accurately the total radiated x-ray yield and power from z-pinch implosion experiments at the Z machine with high accuracy. The Z machine is capable of outputting 2 MJ and 330 TW of x-ray yield and power, and accurately measuring these quantities is imperative. We will describe work over the past several years which include the development of new diagnostics, improvements to existing diagnostics, and implementation of automated data analysis routines. A set of experiments on the Z machine were conducted in which the load and machine configuration were held constant. During this shot series, it was observed that the total z-pinch x-ray emission power determined from the two common techniques for inferring the x-ray power, a Kimfol filtered x-ray diode diagnostic and the total power and energy diagnostic, gave 449 TW and 323 TW, respectively. Our analysis shows the latter to be the more accurate interpretation. More broadly, the comparison demonstrates the necessity to consider spectral response and field of view when inferring x-ray powers from z-pinch sources. © 2014 AIP Publishing LLC.